20 Quick-Check Red User Instructions
Comm Port:
This allows you to choose where to export the "ON PRINT" logged
data.
a. Cell = Setup Cell for data export (4 pin Lemo)
b. Com1 = Not used
c. RADIO = Future development
d. Disabled = to turn off
Timed: This is used when the operator wants to record a live load/pull and
store it internally. Can store up to 255 force readings.
Log Rate
:
The speed at which the logging of force will be recorded. Based
on seconds (Example 5 sec = Unit will record the force once every
5 seconds)
Log Count
:
This is how many logs the operator wants to record. Range is 0 to
255 recorded readings.
Auto Send
: Enable/Disable
Can export the readings live if enabled.
Comm Port
:
This allows you to choose where to export the "ON PRINT" logged
data.
a. Cell = Setup Cell for data export (4 pin Lemo)
b. Com1 = Not used
c. RADIO = Future development
d. Disabled = to turn off
Overload: This is used when an operator or supervisor wants to know if
their equipment is being overloaded. This will be a percentage based on
capacity and will be below 100% of the Quick Check. (This not the same
as the 120% overload that the Quick Check records based on capacity.)
Maximum Records = 255
Threshold %
: (based on capacity)
This the percentage above where the unit will record the peak
load.
Re-arm %
: (based on capacity)
This is the load percentage the force must drop below before the
"OVERLOAD" will reset (re-arm) and store another load.
Holdoff Seconds
:
Once the unit has recorded this overload, how much time will pass
before this unit will re-arm and allow to record the overload again.